IHEP logo
IHEP 2001 - 20 (In Russian)
G.A.Bogdanova, M.V.Vasil'ev, V.Yu.Volkov, A.E.Kiryunin, L.L.Kurchaninov, M.S.Levitsky, A.A.Minaenko, A.M.Moiseev, A.V.Pleskach, E.A.Usenko, Yu.P.Tsyupa
The stand for the measurements of vertex detector silicon counters of SVD set up
Protvino, 2001.-p.8, figs.5, tables 1, refs.:8.


Test bench for the quality check of silicon microstrip detectors is described. The measurements are based on the irradiation of detectors by b-source Sr90. Four detectors have been prepared for the beam run of the SVD setup (April 2000). The results of measurements are presented.


ZIPped PostScript PDF

Preprints'2001 list Support team